SEIKO GROUP DSC ・Sample: Azoxyanisole ・DSC:DSC6200 ・Sample:5mg ・Scan Rate:20゚C/min DSC Prensinhas e Cadinhos Sample Pan 1-7 Sealed Type 8-15 Open Type 16-17 Hermetic Type Panelinhas para alta pressão DSC Modelo DSC 7020 DSC 6100 Tipo de DSC DSC 6220 DSC 6300 Heat Flux DSC Faixa de Temperatura -170 ~ 725°C -170 ~ 500°C -170 ~ 725°C Ambiente ~ 1500°C Ruído Médio 0.1µW 0.1µW 0.2µW 10µW Sensibilidade 0.2µW 0.2µW 0.4µW 20µW Taxa de Aquecimento 0.01 ~ 100°C/min 0.01 ~ 20°C/min 0.01 ~ 100°C/min 0.01 ~ 100°C/min Amostrador Automático Opcional Opcional Opcional Indisponível Controlador de Gás de Purga Opcional Opcional Opcional Opcional Capacidade Máxima: Open Pan 100µL 100µL 100µL 50µL Capacidade Máxima: Hermetic Sealed 15µL 70µL 15µL Capacidade Máxima: Sealed 15µL Sistema Automático de Resfriamento com LN2 (Opcional) -150 ~ 500°C -150 ~ 500°C -170 ~ 725°C Indisponível Sistema Eletrônico de Resfriamento (Opcional) -70 ~ 420°C C Type: -75 ~ 210°C D Type: -30 ~ 250°C C Type: -75 ~ 470°C D Type: -30 ~ 380°C Indisponível Cooling Can (Opcional) -170 ~ 725°C -170 ~ 500°C -170 ~ 725°C Indisponível 15µL DSC DSC DSC OVAL SENSOR Oval Sensor Plate Reference Pan Sample Pan Heater Sub Heat Sink Oval Sensor Plate Heat Filter Main Heat Sink Oval DSC Diagram Main Heat Sink Lead Wire Oval DSC Cross Section ・ High Resolution (Quick responsibility) ・Low thermal resistance by 2nd dimension thermo flow ・The shortest heat flow between sample and reference ・ Protect the sensitivity down (by reducing noise lever) ・ Heat sink substrate used as sensor holder also ・ Filter stored inside sensor in order to cut the noise. DSC • Os DSCs da SII NanoTechnology – Seiko são os únicos DSCs com forno Elipsoidal. • Melhor resolução, alta sensibilidade e excelente planicidade da linha base. Exclusividade da SEIKO SII NanoTechnology DSC Full range Cooling system Electronic Cooling System, Automatic LN2 cooling system Robust Auto-sampler TG/DTA Application of Highway TA Research & Development Application ー DTG(TG) and DSC peak separation. ( for reaction of different activation energies ) ー Data simulation of measurements that are difficult such as extremely high ( ex. 106℃/min ) or extremely low heating rates ( ex. 10-3℃/min ) . Reduced measurement time for Quality Control ー Measurement time → 1/2~1/10 DSC Peak Separation Synthetic Peak Separated Peak Measured Data DSC High and Low Heating Rate Simulation Poliamida 0.00 5ºC/min 10ºC/min 15ºC/min -5.00 20ºC/min 25ºC/min DSC mW -10.00 -15.00 -20.00 120.0 140.0 160.0 180.0 200.0 Temp Cel 220.0 240.0 260.0 280.0