Pág. 1 / 4 Number: Calibration Certificate Serial Number: 0034/09 Sealing Label: 1901509 – VentMeter Analisador de Fluxo 00038 Date: 000063 02/06/2009 General Data Magnamed Tecnologia Médica Addres: R. São Paulino, 221 – Vila Mariana – São Paulo - SP Instrument: VentMeter – Lung Ventilator Analyzer Model: 1901509-01 PL Serial Number: 00038 Flow: 0 – 150 L/min Pressure: -40 – 200 hPa M Range E Client: 1901509-NE-48-01 Calibration Date 00033 00034 02/06/2009 SA Calibration Proceedings Instrument Number Proceedings “The instrument was calibrated as to a Standard of the Quality and Calibration Sector – Magnamed Tecnologia Médica S.A. The standard flow meter was traced together with IPT/CMF – RBC/INMETRO according to certificate N° 93 213-101, valid until 10/03/2011. The standard pressure gauge was traced with ABSi/ RBC-INMETRO, according to certificate N° 60768/09, valid until 26/01/2011.” “The calibration was executed with the instrument in its position use at temperature 18,4 ± 0,5°C, relative humidity of 48,0 ± 1,5% and atmospheric pressure of 1017,7 ± 1,0 hPa, staying for 30 minutes for stability and balance of the environmental parameters.” The presented results have restricted significance and are only applied to the tested sample. The use of this document for promotional purposes as well as its partial or total reproduction depends on the approval in writing of the Magnamed Tecnologia Médica S.A. MAGNAMED Tecnologia Médica S.A. R. São Paulino, 221 – Vila Mariana – São Paulo – SP – Brazil – CEP: 04019-040 Fone: (11) 5081-4115 – Fax (11) 5081-3591 – email: [email protected] CNPJ: 01.298.443/0001-73 - Inscrição Estadual: 149.579.528.111 Pág. 2 / 4 Number: Calibration Certificate Serial Number: 1901509 – VentMeter Analisador de Fluxo 0034/09 Sealing Label: 00038 Date: 000063 02/06/2009 Results 1. Calibration of Pressure Measurements (1mbar = 1hPa ≅ 1 cmH2O) Environmental Conditions – Pressure Measurements: Temperature: 18,4 ± 0,5 °C Atmospheric Pressure: 1017,7 ± 1,0 hPa -20,1 -10,2 -5,1 5,0 10,0 25,2 50,4 75,6 95,6 -20,3 -10,2 -5,1 4,9 10,0 24,8 50,3 75,6 95,6 Mean Pressure (cmH2O) sx Expanded Uncertainty Coverage Factors k -20,4 -10,2 -5,0 4,8 9,8 24,9 50,2 75,5 95,7 -20,3 -10,2 -5,1 4,9 9,9 25,0 50,3 75,5 95,7 0,07 0,03 0,03 0,06 0,06 0,11 0,05 0,06 0,06 0,15 0,06 0,07 0,11 0,11 0,21 0,09 0,11 0,11 2 2 2 2 2 2 2 2 2 PL -20,3 -10,1 -5,0 5,0 9,9 25,1 50,3 75,4 95,8 Relative Humidity: 48,0 ± 1,5% E Measured Pressure (cmH2O) Cicle 1 Cicle 2 ↓ ↑ ↓ ↑ M Indicated Pressure (cmH2O) -20 -10 -5 5 10 25 50 75 95 Date: 02 / 06 / 2009 2. Calibration of Flow Measurements Environmental Conditions – Pressure Measurements: Atmospheric Pressure: 1017,7 ± 1,0 hPa SA Temperature: 18,4 ± 0,5 °C Date: 02 / 06 / 2009 Relative Humidity: 48,0 ± 1,5% Neonate flow Sensor: Indicated Flow (L.min-1) Mean Flow (L.min-1) sx Expanded Uncertainty 0,5 1 5 10 15 20 0,57 0,02 0,05 1,01 0,03 0,07 4,93 0,03 0,06 10,00 0,03 0,06 14,99 0,04 0,07 19,97 0,03 0,05 Coverage Factors k 3,31 2,21 2,04 2 2 2 The presented results have restricted significance and are only applied to the tested sample. The use of this document for promotional purposes as well as its partial or total reproduction depends on the approval in writing of the Magnamed Tecnologia Médica S.A. MAGNAMED Tecnologia Médica S.A. R. São Paulino, 221 – Vila Mariana – São Paulo – SP – Brazil – CEP: 04019-040 Fone: (11) 5081-4115 – Fax (11) 5081-3591 – email: [email protected] CNPJ: 01.298.443/0001-73 - Inscrição Estadual: 149.579.528.111 Pág. 3 / 4 Number: Calibration Certificate Serial Number: 1901509 – VentMeter Analisador de Fluxo 0034/09 Sealing Label: 00038 Date: 000063 02/06/2009 Infant Flow Sensor: Mean Flow (L.min-1) sx Expanded Uncertainty 1 5 10 20 30 40 50 1,28 0,12 0,26 5,20 0,12 0,24 10,23 0,08 0,16 19,98 0,08 0,16 30,17 0,10 0,20 39,98 0,14 0,28 50,31 0,15 0,30 5 10 25 50 75 100 125 150 sx Expanded Uncertainty 5,26 0,26 0,52 10,11 0,10 0,21 25,16 0,27 0,53 50,16 0,16 0,31 75,20 0,44 0,88 99,30 0,36 0,72 125,10 0,61 1,22 150,77 0,44 0,88 SA where: Mean Flow (L.min-1) M Indicated Flow (L.min-1) PL Adult Flow Sensor: The experimental standard deviation of the mean ( sx ) was calculated by: sx = Standard Deviation Coverage Factors k 2,21 2,04 2,01 2 2 2 2 E Indicated Flow (L.min-1) sx = Coverage Factors k 2,04 2,01 2 2 2 2 2 2 sx n n = repeated measurements The expanded uncertainty was calculated by standard uncertainty multiplied coverage factors k. The assigned expanded uncertainty corresponds a coverage probability of approximately 95% Standards Flow Meter: TSI Flowmeter Calibrador Padrão de Fluxo Magnamed, id n. 00033, serial number 40400842004, calibration certificate n° 93 213-101, de 10.03.2009, traced together with IPT/CMF – RBC/INMETRO, valid until 10/03/2011. Pressure Meter:: Digital Pressure Gauge 0 - 1600MMCA, id n. 00034, serial number 59601, calibration certificate n° 60768/09, de 26.01.2009, traced with ABSi/ RBC-INMETRO, valid until 26/01/2011. The presented results have restricted significance and are only applied to the tested sample. The use of this document for promotional purposes as well as its partial or total reproduction depends on the approval in writing of the Magnamed Tecnologia Médica S.A. MAGNAMED Tecnologia Médica S.A. R. São Paulino, 221 – Vila Mariana – São Paulo – SP – Brazil – CEP: 04019-040 Fone: (11) 5081-4115 – Fax (11) 5081-3591 – email: [email protected] CNPJ: 01.298.443/0001-73 - Inscrição Estadual: 149.579.528.111 Pág. 4 / 4 Number: Calibration Certificate Serial Number: 1901509 – VentMeter Analisador de Fluxo 0034/09 Sealing Label: 00038 Date: 000063 02/06/2009 Calibration Procedures The VentMeter was calibrated according internal proceedings 1901509-NE-48-01 – Instruction of Calibration (review 02). The method is based on comparation of VentMeter indicated value and the Standard indicated value measurements SA M PL E The flow results presented in this certificate represents the mean of three repeated measurements. São Paulo, June 9, 2009 ____________________________________ Toru Miyagi Kinjo Eletronic Engineer Technical Responsal CREA n° 5061555031 The presented results have restricted significance and are only applied to the tested sample. The use of this document for promotional purposes as well as its partial or total reproduction depends on the approval in writing of the Magnamed Tecnologia Médica S.A. MAGNAMED Tecnologia Médica S.A. R. São Paulino, 221 – Vila Mariana – São Paulo – SP – Brazil – CEP: 04019-040 Fone: (11) 5081-4115 – Fax (11) 5081-3591 – email: [email protected] CNPJ: 01.298.443/0001-73 - Inscrição Estadual: 149.579.528.111